SP 2000UV/UVPC Visible Spectrophotometer



Can you believe? An UV-Vis spectrophotometer costs less than two thousand, but with power on self-testing, automatic wavelength calibration, light source auto- switching, PC controlled, etc. In the past, it might cost several two-thousand even decuple with these advanced functions.
SP-2000UV UV-Vis spectrophotometer takes use of high precision sine mechanism as a wavelength driver. Program control, wavelength accuracy, and the linearity over the full wavelength are obviously better than those of the congener instruments in which gear and rack is used as a wavelength driver, and a gauge dial is used for wavelength display.

Specification:

MODEL
SP-2000UV
SP-2000UVPC
Optic system
Littrow type optic system with 1200 grooves/mm diffraction grating monochromator.
Light path
100mm
Spectrum band width
5nm
Wavelength range
200-1000nm
190-1100nm
Wavelength accuracy
±2nm (automatic system calibration )
Wavelength repeatability
1.0nm
Stray light
≤0.5%T (at 220nm & 340nm)
Photometric mode
T, A, and C measurement;
Photometric range
0-125.0%T, -0.097-3.000A, 0-9999C, 0-9999F
Photometric accuracy
±0.5%T
Photometric repeatability
0.3%T
Stability
±0.003A/hr (at 500nm, after warming up for 1 hour)
Baseline flatness
±0.004A
 Main Specification:
Display 2*20 characters LCD with blacklight function
Light Source 6V10W Halogen Lamp(2000 hours)
Deuterium lamp(1000 hours)
Sample Compartment100mm Optical Path
Standard Cell Holder 10mm
Standard Cell Configuration 10mm Glass cell(4pcs)
Power Requirement 110/220VAC
 Basic testing mode:
Basic measurement Photometric measurement,(A/T)
Concentration measurement(C )
Concentration measurement Slope method (C=K*A)
Standard curve method(C=K*A+b)
Software functions Win-spec workstation, SP-2.33EN
Photometric analysis T, A, and C measurement
Quantitative analysis Standard curve measurement
Kinetic analysis: measurement with time-scanning
Spectrum scan analysis spectrum scan function
 Basic instrument functions:
100%T or 0%T setup Automatic.
System self-testing Wavelength self-testing, light soures self-testing,filter switching,energy calibrating
Light source switching The switching point can be set at any position at the wavelength range of 325-355nm to improve testing accuracy
Light source management Light source on/off status can be controlled to elongate its working life
Signal output RS-232C, Analog
Communication function Intercommunication can be realized after connecting to PC
 Extended function (Accessories or workstation software should be chosen.)
Automatic multi-cells holder Auto-sorting or manual-sorting test, 4 cells and 6 cells for choice.
Micro-volume testing 100μl, as the minimum volume. (Special option cell holder and cuvette should be used.)
Constant temperature testing Testing samples under the condition of constant temperature. This function is especially suit for kinetic analysis. Electronic constant temperature accessory or water cycling constant temperature accessory should be chosen.
USB communication USB Port for choice, which fits the developmental tide of modern computer.
 
Spectral scanning
In the instrument's wavelength range, run spectra scanning for sample according to the setting wavelength range, get the curve of Absorption/Transmittance vs. wavelength to analyze the spectrum characteristic and identify the maximum absorbance peak. User can do spectrum smoothing, peak or valley detecting, and overlap several spectrum curves to do four basic calculations, print and save the data.
 
Note: There may not be special notifications when the technical specification,profile and configuration changed. Please, the products will be in agreement with the real models.