SP2000 Visible Spectrophotometer



Recommended application:

SP-2000 is a visible spectrophotometer special designed for micro amount sample. The sample compartment can hold cuvettes with100mm light path, and diversified cuvettes with light path at a range of 5-50mm also can be hold. SP-2000 is used as a teaching instrument in basis laboratories, which is used in colleges for specialty of environmental protection. It's also used as an instrument for products quality control at sugar-refinery, food and drink industry. It's better used as a monitoring device that controls the let of waste water in industries, such as electricity generating, water supply, paper making, plating, printing and dyeing, and so on.
Main features:

* The large sample compartment with long path detecting system, enable diversified cuvettes be hold at the range of 5-100mm.
* Novel designed auto-collimated light path with grating monochromatic, guarantees excellent optical performance.
* The SCM technology enables the instrument to be intelligent and digital.
* Function of 0%T and 100%T self-adjusting, which enables the operation simple and human.
* Large display with LCD, which enables the testing parameters, be displayed as transmittance, absorbance, and concentration.
* High precision of data processing that reaches one percent of a hundred thousand, which makes the transform between transmittance and absorbance with little error be possible.
* Filters auto-switching at the whole wavelength range, enables the instrument without additional manual operation.
* Switching power supply, with self-adaptive inputting voltage of 85~264VAC, (that is abbreviated to “SMPS”), guarantees that the instrument can work normally with different power supply at any environment.

 
Main Specification:
Display 3.5bits digital LCD Displayer
Light Source 6V10W Halogen Lamp(2000 hours)
Sample Compartment(Max) 50mm Optical Path
Standard Cell Holder 50mm
Standard Cell Configuration 10mm Glass cell(4pcs)
Power Requirement 110/220VAC,50/60Hz,±10%
 
Basic testing mode:
Basic measurement Photometric measurement,(A/T)
Concentration measurement (C )
Concentration measurement Slope method (C=K*A)
Standard curve method
Software functions Win-spec workstation, SP-1.33EN
Photometric analysis T, A, and C measurement
Quantitative analysis Standard curve measurement;
Kinetic analysis Measurement with time-scanning.
 
Basic instrument functions:
100%T or 0%T setup Automatic.
Signal output RS-232C, Analog
Communication function Intercommunication can be realized after connecting to PC
 
Photometric Performance:
Photometric Analysis T, A, and C measurement;
Photometric Range 0-125.0%T,0-1.999A, 0-1999C, 0-1999F
Photometric Accuracy ±0.5%T
Photometric repeatbility 0.3%T
Stability ±0.003A/hr (at 500nm, after warming up for 1 hour)
 
Optical Performance:
Optic System Littrow type optic system with 1200 grooves/mm diffraction grating monochromator
Light path 100mm
Spectrum bandwidth 5nm
Wavelength range 325-1000nm
Wavelength accuracy ±2nm
Wavelength repeatability 1nm
Stray light ≤0.5%T(at 360nm )
 
 
Standard configuration
50mm manual four-cell holder, 1pcs
10*10*45 mm standard glass cuvettes, 4pcs
RS-232C connection cord, 1pcs
Power cable, 1pcs
User manual, 1pcs
Operation procedure, 1pcs
Win-spec software SP-1.33EN, 1pcs

Function configuration
Standard type: SP-2000
The operation can be carried out by the instrument. It also can be controlled by PC, which is easy to master, and plenty of testing data can be stored, data processing such as calculation, statistic analysis and tab also can be realized.

Note: There may not be special notifications when the technical specification,profile and configuration changed. Please, the products will be in agreement with the real models.